MultiSystem 4010
Operating Instructions
© Hydrotechnik GmbH • All rights reserved
Rev. 1.5 / 120816 english
Page 33 of 45
Highlight a channel, press
, select the desired colors and confirm with ;
select the desired symbol and confirm with
; repeat this for all desired channels.
Press
to enable/disable the colors, or to enable/disable the symbols;
press
to confirm the settings.
Press
to terminate the settings in the display menu.
6.6. Program a Recording
Highlight „Memory“ in the main menu and press :
In this menu you define, which channels are recorded and how:
Channels highlight the function and press
to select the channels to be recorded:
Highlight a channel and press
to check / uncheck it; all channels with green
check marks will be recorded; press
to finish the channel selection
Storing time defines how long the recording shall last; first enter a value and then select the
time unit second, minute or hour
Scan rate defines the time intervals between two measured values; first enter a value and
then select the time unit milli-second, second or minute
Trigger event by defining a trigger you may start the recording at exactly that moment when
interesting data are measured; one of the channels Ch1 … Ch7 can be supervised
for the fulfillment of a condition; with the trigger event “Key” you may start the