Setting emission degree according to the table
The emission degree can be set via an integrated material table for common metals.
Select the function “Tab” via the button “F2”. A materials table is displayed.
Use the function buttons “F1” and “F3” to set the present material. The selected material is marked with a bar. The
degree of emission changes accordingly.
“Default” = Emission value of most materials (0.95)
“Ox Aluminum” = Aluminium, oxidised (0.30)
“Ox Brass” = Brass, oxidised (0.50)
“Ox Copper” = Copper, oxidised (0.60)
“Paint” = Painted surface (0.93)
Conrm the selection with “OK” (button “F2”). To terminate the setting, push the measuring button (5) or the button
“MODE”.
Below the technical data, you will nd a table with typical materials and their emission degrees.
Many organic materials have an emission degree of 0.95. Therefore, the factory setting of the emission degree
is 0.95.

The meter is equipped with a continuous measurement function for long-term measurements.
Switch on the meter by briey pushing the measuring button (5).
Select the setting menu by pushing the button “MODE” within the pre-set deactivation time
The continuous measurement is switched on and off via the function button “F1”. Every push will switch the function
on or off.
Continuous measurement off
Continuous measurement on